Determination of calcium in water by flow injection analysis — inductively coupled plasma (FIA-ICP) emission spectrometry |
| |
Authors: | Phillip L Kempster Jacobus F Staden and Henk R Vliet |
| |
Institution: | (1) Department of Water Affairs, Hydrological Research Institute, Private Bag X313, 0001 Pretoria, South Africa;(2) Department of Chemistry, University of Pretoria, 0002 Pretoria, South Africa |
| |
Abstract: | Summary Using an 11 ml centrifugal type spray chamber, a rapid sample introduction rate of 320 injections per hour may be achieved when interfacing flow injection analysis and inductively coupled plasma emission spectrometric techniques. Calcium was determined in water samples with a relative standard deviation of better than 3.5% over the analytical range of 10 to 200 mg · l–1 Ca. Mean recovery was 97%.
Calciumbestimmung im Wasser mit dem Flow-injection-Prinzip kombiniert mit der Induktions-gekoppelten Plasma-Emissions-Spektrometrie (FIA-ICP) |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|