Abstract:Correction equations of the coincidence-summing effect for efficiencies of HpGe detector based on the decay scheme were developed by considering the summing up to triple coincidence. The correction equations which do not dependent on the kind of the Ge detector are very useful for efficiency calibrations of a Ge detector in the energy region from 60 to 400 keV by using75Se radionuclide even with very short source-to-detector distances.