Isocratic high-performance liquid chromatographic determination of plasma adenosine
Authors:É. Huszár  E. Barát  M. Kollai
Affiliation:(1) Department of Pathophysiology, National Korányi Institute for TB and Pulmonology, 1529 Budapest, Hungary;(2) Present address: Pihenö u. 1, H-1529 Budapest
Abstract:Summary Due to manifold physiological and cardioprotective actions of adenosine, the demand for a simple but accurate method to determine its concentration in plasma is increasing. The aim of this study was firstly to develop a simple isocratic method instead of the gradient elution or peak-shifting techniques used earlier and secondly to check conflicting data on the composition of ldquostop-solutionrdquo, added to the sample in order to prevent changes in adenosine concentration. Isocratic elution improved signal to noise ratio and concentrations of 100 mgrmol L–1 dipyridamole and 2.5 mgrmol L–1 erythro-9(2-hydroxy-3-nonyl)adenine in the blood sample effectively prevented both adenosine formation and degradation, even without the use of a 5prime-ecto-nucleotidase inhibitor. Lowering the concentration of dipyridamole to 25 mgrmol L–1 caused more than a tenfold increase of adenosine concentration in two out of five cases and even 100 mgrmol L–1 dipyridamole alone is not sufficient to inhibit adenosine deaminase in blood samples.
Keywords:Column liquid chromatography  Adenosine in blood  Isocratic elution  Stop-solution