首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Cholesterol biosensor based on amino-undecanethiol self-assembled monolayer using surface plasmon resonance technique
Authors:Solanki Pratima R  Arya Sunil K  Nishimura Y  Iwamoto M  Malhotra B D
Institution:Biomolecular Electronics and Conducting Polymer Research Group, National Physical Laboratory, Dr. K. S. Krishnan Marg, New Delhi-110012, India.
Abstract:Cholesterol oxidase has been covalently immobilized onto 11-amino-1-undecanethiol hydrochloride (AUT) self-assembled monolayer (SAM) fabricated on gold (Au) substrates using glutaraldehyde as a cross-linker. These ChOx/AUT/Au bioelectrodes characterized using contact angle (CA) measurements; electrochemical technique and atomic force microscopy (AFM) have been utilized for the estimation of cholesterol in solution using the surface plasmon resonance (SPR) technique. These biosensing electrodes exhibiting linearity from 50 to 500 mg/dL of cholesterol in solution and sensitivity of 1.23 m0/(mg dL), can be used more than 20 times and have a shelf life of about 10 weeks when stored at 4 degrees C.
Keywords:
本文献已被 PubMed 等数据库收录!