首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Flow-through sensor based on derivative synchronous fluorescence spectrometry for the simultaneous determination of pyrene, benzo(e)pyrene and benzo(ghi)perylene in water
Authors:P Cañizares and M D Luque de Castro
Institution:(1) Department of Analytical Chemistry, Faculty of Sciences, University of Córdoba, E-14004 Córdoba, Spain
Abstract:A flow-through/first derivative synchronous spectrofluorimetric sensor for the determination of PAH has been described. This sensor has been used for the simultaneous determination of PAH mixtures (pyrene, benzo(e)pyrene and benzo(ghi)perylene). Linear calibration ranges between 10 and 500 ng/ml with acceptable precision (repeatability, expressed as relative standard deviation, smaller than 4.6%, and sampling frequency of 12 h–1) have been obtained. The method has been applied to the determination of the target analytes in spiked water samples with excellent results (recoveries between 94 and 108%).
Keywords:
本文献已被 SpringerLink 等数据库收录!