Candidate reference method: Determination of valproic acid in serum by isotope-dilution mass spectrometry |
| |
Authors: | F Susanto H Reinauer |
| |
Institution: | (1) Department of Clinical Biochemistry, Diabetes Research Institute, Geinrich-Heine-University, Auf'm Hennekamp 65, D-40225 Düsseldorf, Germany |
| |
Abstract: | Summary A method has been developed for the determination of valproic acid, without derivatization, in human serum by isotope-dilution
mass spectrometry using labelled 2-propyl3,3,3-d3]valeric5,5,5-d3] acid as internal standard for accurate quantification of the concentration of valproic acid in the sample. After acidification,
the analyte and internal standard are extracted withn-hexane. The amounts of valproic acid in the serum are calculated from the isotope ratio of valproic acid to labelled valproic
acid, which is measured by electron impact (EI) and chemical ionization (CI) selected ion monitoring (SIM).
The concentrations of valproic acid in sera measured using isotope-dilution mass spectrometry are compared with results from
gas-liquid chromatography (GLC) and fluorescence polarization immunoassay (FPIA). The accuracy, precision and recovery of
the GC-MS methods are discussed. The coefficient of variation determined from duplicate samples was less than 1.5%. The detection
limit was 10 ng mL−1 at a signal-to-noise ratio of 3∶1.
Part of this work was presented at the Kongre? der Deutschen Gesellschaft für Laboratoriumsmedizin, Berlin, 1994. |
| |
Keywords: | Gas chromatography-mass spectrometry Isotope-dilution mass spectrometry (IDMS) Valproic acid in serum |