Abstract: | A simple and rapid ultra‐high‐performance liquid chromatographic (UHPLC) for the simultaneous determination of meropenem and ciprofloxacin in human plasma was developed and validated. All of the analytes were separated in <5 min. A solid‐phase extraction method was applied from sample preparation. Analytical separation was performed on a Poroshell SB C18 column (50 × 2.1 mm, 2.7 μm particle size) with photodiode array (PDA) detection. Meropenem and ciprofloxacin were determined at wavelengths of 300 and 277 nm, respectively. The mobile phase was a mixture of acetonitrile–10 mm ammonium acetate–methanol in gradient elution. The method has been validated for both drugs in gastric surgery for cancer patients. The method showed good linearity with correlation coefficients, r2 = 0.994 for the two drugs, as well as high precision (RSD < 10.5% in each case); accuracy ranged from ?5.8 to +6.0%. The limit of quantitation of the two drugs was established at 0.02 and 0.01 μg/mL, respectively. Meropenem, ciprofloxacin and the internal standard were extracted from human plasma with a mean recovery ranging from 92.5 to 98.6%. The method was applied to quantify the drugs dosage in complicated gastric surgery patients. |