Determination of Meloxicam Using Europium Sensitized Luminescence in the Presence of Co-Luminescence Reagents |
| |
Authors: | Salma Muhamed Z Al-Kindy Salim K Al-Habsy Fakhr Eldin O Suliman Haidar A J Al-Lawati |
| |
Institution: | (1) Department of Chemistry, College of Science, Sultan Qaboos University, Al-Khod 123, Box 36, Muscat, Sultanate of Oman |
| |
Abstract: | A sensitive time- resolved luminescence method for the determination of meloxicam (MX) in methanol and in aqueous solution
is described. The method is based on the luminescence sensitization of europium (Eu3+) by formation of ternary complex with MX in the presence of 1,10- phenanthroline as coligand, Tween-80 as surfactant and
gadolinium ion as a co-luminescence reagent. The signal for Eu- MX-1, 10- phenanthroline is monitored at λex = 360 nm and λem = 620 nm. Optimum conditions for the formation of the complex in aqueous system were 0.01 M TRIS buffer, pH 8.0, 1,10- phenanthroline
(6.0 × 10−6 M) , Gd3+ (7.0 × 10−6 M), Tween-80 (0.28%) and 1.75 mM of Eu3+ which allows the determination of 20–800 ppb of MX with limit of detection (LOD) of 7 ppb. The relative standard deviations
of the method range between 0.1 and 1.1% indicating excellent reproducibility of the method. The proposed method was successfully
applied for the assay of MX in pharmaceutical formulations, plasma and in urine samples. Average recoveries of 99.8 ± 1.1%,
100.2 ± 0.9% and 100.9 ± 1.1% were obtained for MX in tablet, plasma and urine sample respectively. |
| |
Keywords: | |
本文献已被 PubMed SpringerLink 等数据库收录! |
|