Analytical performance of glucometers in a tertiary care hospital
Authors:Angela Somavilla Higioka  Renata Cristina Messores Rudolf-Oliveira  Flavia Martinello  Maria Luiza Bazzo
Institution:1.Residência Multidisciplinar em Saúde, University Hospital,Federal University of Santa Catarina,Florianópolis,Brazil;2.Laboratório de Pesquisa III (segundo andar), Divis?o de Análises Clínicas,Hospital Universitário Professor Polydoro Ernani de S?o Thiago - Universidade Federal de Santa Catarina,Florianópolis,Brazil;3.Program of Post Graduation in Pharmacy, Department of Clinical Analysis,Federal University of Santa Catarina,Florianópolis,Brazil;4.Departamento de Análises Clínicas, Centro de Ciências da Saúde,Universidade Federal de Santa Catarina,Florianópolis,Brazil
Abstract:Point-of-care (POC) testing of glucose (glucometers) represents a convenient alternative to monitor glycemia since the measurement procedure is performed without delay after sampling of the capillary blood, thereby avoiding the metabolism by the blood cells of glucose present in plasma. Likely because of sample instability, there is no proficiency test provider in Brazil for this type of POC sample. In this context, this study aimed to evaluate the analytical performance of glucometers used in a tertiary care hospital. The glucometers used were the Accu-Chek Performa® model from Roche Diagnostics, which use the principle of amperometry. The reference method was the reaction with modified hexokinase/glucose-6-phosphate in a Dimension® device. The stability evaluation of the control samples showed that it can be performed up to 90 min after the collection of whole blood samples. In the two rounds performed, only one result of the 17 glucometers evaluated was out of the threshold of two standard deviation. Thus, this method for control of glucometers met the expectations and enabled comparing the glucometers in a hospital. Given the current quality guidelines, daily internal quality control of glucometers is recommended, besides at least two annual comparisons between the results of the glucometers and the reference method and one EQA every 3 months.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号