首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Nanomolar detection with high sensitivity microfluidic absorption cells manufactured in tinted PMMA for chemical analysis
Authors:Floquet Cedric F A  Sieben Vincent J  Milani Ambra  Joly Etienne P  Ogilvie Iain R G  Morgan Hywel  Mowlem Matthew C
Institution:a National Oceanography Centre, University of Southampton, Southampton SO14 3ZH, UK1
b Nanoscale Systems Integration Group, University of Southampton, Southampton SO17 1BJ, UK2
Abstract:We describe a novel, cost effective and simple technique for the manufacture of high sensitivity absorption cells for microfluidic analytical systems. The cells are made from tinted polymethyl methacrylate (PMMA) in which microfluidic channels are fabricated. Two windows (typically 250 μm thick, resulting in little optical power loss) are formed at either end of the channel through which light is coupled. Unwanted stray light from the emitter passes through a greater thickness of the tinted substrate (typically the length of the cell) and is preferentially absorbed. In effect, this creates a pin-hole configuration over the length of the absorption cell, providing improved performances (sensitivity, S/N ratios, baseline noise and limit of detection) when used as an absorption cell compared to clear substrates. The method is used to achieve a LOD of 20 nM with a colourimetric iron assay and a LOD of 0.22 milli-absorption units with a pH assay.
Keywords:Microfluidic  PMMA  Colourimetric assays  Absorption  Microsensors
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏