Potentiometric immunosensor based on antiserum of Japanese B encephalitis immobilized in nano-Au/polymerized o-phenylenediamine film
Authors:Lingyan Zhang  Ruo Yuan  Xiaoqing Huang  Yaqing Chai  Shurui Cao
Institution:aChongqing Key Laboratory of Analytical Chemistry, College of Chemistry and Chemical Engineering, Southwest China Normal University, Chongqing 400715, China;bSchool of Material Science and Chemical Engineering, Tianjin Polytechnic University, Tianjin 300160, China
Abstract:A novel potentiometric immunosensor for detection of Japanese B encephalitis vaccine was developed by immobilizing antiserum of Japanese B encephalitis on nano-Au/polymerized o-phenylenediamine (o-PDA) film on the platinum (Pt) electrode. The performance and factors influencing the performance of the resulting immunosensor were studied. The immunosensor showed a specific response to Japanese B encephalitis vaccine in the range 1.1 × 10−8 to 2.0 × 10−6 lgpfu/ml (plaque forming unit) with a detection limit of 6 × 10−9 lgpfu/ml. The correlation coefficient is 0.9986. The incubation time, incubation temperature, pH, reproducibility and stability of the immunosensor were also studied. The present work supplied a promising test method for biological products.
Keywords:Potentiometric  Nano-Au  Japanese B encephalitis vaccine  Immunosensor