Potentiometric response of a commercial (Ag2S+PbS)-membrane selective electrode to Hg(II) in aqueous solution |
| |
Authors: | Nj. Radić |
| |
Affiliation: | (1) Laboratory for Analytical Chemistry, Faculty of Technology, University of Split, Split, Yugoslavia |
| |
Abstract: | Summary The preparation and performance of an Hg(II)-ISE are described. The analytical behaviour of this electrode, made by chemical treatment of a commercial silver sulphide type Pb(II)-ISE, is discussed in terms of potentialvs. concentration curves, potentialvs. pH curves, and selectivity. The response to fixed cHg(II) as the pH is increased in the alkaline region corresponds to decrease in the free Hg2+concentration down to pHg 17. The estimated limit of detection of total Hg(II) concentration is 10–5M.
Potentiometrisches Verhalten einer handelsüblichen (Ag2S+PbS)-Membran-Elektrode gegenüber Hg(tt) in wäßriger Lösung Zusammenfassung Herstellung und Wirkungsweise einer gegen Hg(II) ionenselektiven Elektrode wurde beschrieben. Das analytische Verhalten dieser Elektrode, die sich durch chemische Behandlung einer handelsüblichen Silbersulfidelektrode, die für Pb(II)-Ionen selektiv ist, herstellen läßt, wurde an Hand von Potential-Konzentrationskurven, Potential-pH-Kurven und ihrer Selektivität diskutiert. Das Verhalten gegenüber fixemcHg(ii)bei Anstieg des pH ins alkalische Gebiet korrespondiert mit der Abnahme der Konzentration an freien Hg2+-ionen bis herab zu pHg 17. Die Nachweisgrenze liegt schätzungsweise bei einer totalen Hg(II)-Konzentration von 10–5M.
Presented at the Fifth European Conference on Analytical Chemistry (EUROANALYSIS-V), Cracow, Poland, 26–31 August, 1984. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|