Improvements and new capabilities for the multiple Knudsen cell device used in high-temperature mass spectrometry |
| |
Authors: | Heyrman M Chatillon C Collas H Chemin J-L |
| |
Affiliation: | Laboratoire de Thermodynamique et Physico-Chimie Métallurgiques (CNRS UMR 5614, UJF/INPG), ENSEEG-Domaine Universitaire, B.P. 75, 38402 Saint Martin d'Hères, France. mheyrman@ltpcm.inpg.fr |
| |
Abstract: | The thermodynamic properties of condensed phases, i.e. the activities of components, can be determined from partial pressures measured by the Knudsen cell mass spectrometric method. Improvements in accuracy and yield of this method are obtained with the use of twin cells, an idea proposed in the 1960s. The multiple cell method was perfected in 1977 in our laboratory. Changes to molecular beam sampling and furnace assemblies were required to make the multiple Knudsen cell technique work properly. This paper summarizes these prerequisites, and presents a new device and the associated method of measurements, as well as the necessary tests performed with a silver sample in each cell. |
| |