Design and application of a liquid-flow adsorption micro calorimeter based on the Calvet-type DAK 1-1 microcalorimeter (USSR) |
| |
Authors: | J. Seidel |
| |
Affiliation: | (1) Research Institute of Mineral Processing, Academy of Sciences of the GDR, Str. des Friedens 40, 9200 Freiberg, Germany |
| |
Abstract: | The construction and the properties of a liquid-flow adsorption microcalorimeter are described. The calorimeter is based on the commercially available Calvet-type DAK 1-1 microcalorimeter (USSR), which has been equipped with a flow system consisting of a pump, several sets of heat exchangers and an adsorption cell. The minimum detectable heat effect is 1W up to a flow rate of 30 cm3/h.Results on the application of the calorimeter to adsorption and desorption processes of nonionic (Triton X-100) and anionic (styrene phosphonic acid) surfactant solutions in water at an-Al2O3 surface are presented and discussed.
Zusammenfassung Konstruktion und Eigenschaften eines Flüssigkeits-Strömungs-Adsorptions-Mikrokalorimeters werden beschrieben. Das Kalorimeter wurde weiterentwickelt aus dem Calvet-Mikrokalorimeter DAK 1-1 (USSR), das mit einem Strömungssystem aus einer Pumpe, verschiedenen Wärmeaustauschern und einer Adsorptionszelle ausgestattet wurde. Der kleinste nachweisbare Wärmeeffekt liegt bei 1W bis zu einer Durchflussgeschwindigkeit von 30 cm3 h–1. Mit dem Kalorimeter erhaltene Messergebnisse zur Adsorption und Desorption wässeriger Lösungen von nichtionischen (Triton X-100) und anionischen (Styrenphosphonsäuren) Tensiden an-Al2O3 werden dargestellt und diskutiert.
, 1-1 (), , , . 1 30 3/. ( X–100) (- ) -l23. |
| |