首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Electrochemical Study of 4-Nitrophenol at a Modified Carbon Paste Electrode
Authors:María del Mar Cordero-Rando  Manuel Barea-Zamora  Juan M Barberá-Salvador  Ignacio Naranjo-Rodríguez  Juan A Muñoz-Leyva  José L Hidalgo-Hidalgo de Cisneros
Institution:(1) Department of Analytical Chemistry, Faculty of Sciences, University of Cádiz, 11510 Puerto Real, Cádiz, Spain, ES
Abstract: A zeolite-modified carbon paste electrode (CPE) has been used for the determination of 4-nitrophenol by differential pulse voltammetry (DPV). The electrochemical reduction of 4-nitrophenol at −1.0 V is carried out in a Britton-Robinson medium at pH 3.5. The cyclic voltammetric (CV) behaviour has been investigated to study the nature of the process. Studies on the effect of pH were carried out over the pH range 2–9 with the Britton-Robinson buffer solution, and the influence of pH on peak height and peak potential was analyzed. A linear relationship between peak intensity and concentration is obtained in the range 0.2–10 mg L−1, with a detection limit of 0.04 mg L−1; a relative standard deviation of 1.5% for a 5 mg L−1 4-nitrophenol concentration and a relative error of 2.6% were also obtained (n=11). Received March 3, 1998. Revision December 10, 1998.
Keywords::   4-Nitrophenol  differential pulse voltammetry  zeolite-modified carbon paste electrode  
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏