(1) Fachbereich Anorganische Chemie und Kernchemie der Technischen Hochschule Darmstadt, D-6100 Darmstadt, Bundesrepublik Deutschland
Abstract:
Summary Ion-exchange chromatography on a AS4A-column followed by a hollow fiber suppressor is suitable for the separation of sulphite, sulphate, thiosulphate and thiocyanate. Injecting a volume of 110 l, conductivity measurement yields detection limits between 15 g/l for sulphate and 75 g/l for sulphite.Ion-pair chromatography on a LiChrospher-100CH column using tetrabutylammonium ions as lipophilic reagent additionally makes possible the determination of dithionate, peroxodisulphate and tetrathionate, thus including seven anions in one run. Using a micro membrane suppressor and conductivity detection, detection limits between 10 g/l for sulphate and 70 g/l for tetrathionate are achieved with injection of 50 l samples.Herrn Prof. Dr. W. Fresenius aus Anla\ seines 75. Geburtstages gewidmet