A gamma-ray detector for capillary zone electrophoresis and its use in the analysis of some radiopharmaceuticals
Authors:K D Altria  C F Simpson  A K Bharij  A E Theobald
Institution:Department of Chemistry, Birkbeck College, London, England.
Abstract:The construction and evaluation of a gamma-ray detector for capillary zone electrophoresis is described. The detector was shown to have a linear response from the limit of detection 10 to 500 Bqcm-3 corresponding to 5.1 x 10(-17) to 2.55 x 10(-15) gcm-3 43Tc99m. The application of the detector for the analysis of some radiopharmaceuticals is presented.