Abstract: | The high dose standards and dissemination system of electron beams are being established at NIM. The graphite and/ or water calorimeters and liquid chemical dosimeter are to be accepted as standards. The transfer dosimeter selected are alanine/ESR dosimeter and radiochromic film (FWT - 60). Several kinds of radiochromic films, undyed cellulose triacetate, polyethylene and blue cellophane will be recommended as working dosimeter. A series of intercomparison studies are conducted between calorimeter and dichromate dosimeter. Agreement is found within 2%. Water calorimeters and dichromate dosimeters are used to make absolute dosimetric measurements of electron beams. These calibrated beams are then used to calibrate several types of dosimeters, such as alanine, radiochromic films, undyed and dyed polyethylene. Preliminary studies show that water calorimeter and dichromate dosimeter are reproducible and sufficiently accurate for electron beam calibration. The estimated overall uncertainty of the measurement is better than 5% at 95% confidence level. |