首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A voltammetric method for the quantitative determination of midecamycin compared to its simultaneous HPLC determination
Authors:K M Drljevi?-Djuri?  M L Avramov Ivi?  S D Petrovi?  D ? Mijin  M B Jadranin
Institution:1.Medicines and Medical Devices Agency of Serbia,Belgrade,Serbia;2.ICTM—Institute of Electrochemistry,University of Belgrade,Belgrade,Serbia;3.Faculty of Technology and Metallurgy,University of Belgrade,Belgrade,Serbia;4.Hemofarm, Stada, Pharmaceutical and Chemical Industry,Vr?ac,Serbia;5.Institute of Chemistry, Technology and Metallurgy,University of Belgrade,Belgrade,Serbia
Abstract:The aim of this study was to present the quantitative determination of midecamycin at a gold electrode in 0.05 M NaHCO3 using cyclic linear sweep voltammetry. It was found that the value of the oxidative peak of pure midecamycin at 0.85 V vs. SCE at a scan rate of 50 mV s−1 is a linear function of the concentration in the range 0.1693–0.3289 mg cm−3. The value of its reductive peak at 0.3 V vs. SCE is a linear function of the concentration in the range 0.11396–0.3802 mg cm−3. HPLC analysis of the bulk of electrolyte confirmed the data obtained by analysis of the current peak values concerning the correlation with each investigated concentration of midecamycin. By MS spectrometry no degradation products were found in electrolyte during its quantitative determination.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏