首页 | 本学科首页   官方微博 | 高级检索  
     


Study of the thermal stability of Zerodur glass ceramics suitable for a scanning probe microscope frame
Authors:Ondřej Číp  Radek Šmíd  Martin Čížek  Zdeněk Buchta  Josef Lazar
Affiliation:1.Institute of Scientific Instruments of the ASCR,v.v.i. Academy of Sciences of the Czech Republic,Brno,Czech Republic
Abstract:The work presents measurements of the length stability of Zerodur glass ceramic with temperature change. Measurement of this thermal characteristic is necessary for determination of the optimal temperature at which the Zerodur glass ceramic has a coefficient of thermal expansion close to zero. The principle of the measurement is to monitor the length changes using an optical resonator with a cavity mirror spacer made from the Zerodur material to be studied. The resonator is placed inside a vacuum chamber with a temperature control. A tunable laser diode is locked to a certain optical mode of the resonator to monitor the optical frequency of this mode. A beat-note signal from optical mixing between the laser and a stabilized femtosecond frequency comb is detected and processed. The temperature dependence of the glass ceramics was determined and analyzed. The resolution of the length measurement of the experimental set-up is on the order of 0.1 nm.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号