首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Analysis of bromate in drinking water using liquid chromatography-tandem mass spectrometry without sample pretreatment
Authors:Kosaka Koji  Asami Mari  Takei Kanako  Akiba Michihiro
Institution:Water Management Section, Department of Environmental Health, National Institute of Public Health, Wako, Saitama, Japan. kosaka@niph.go.jp
Abstract:An analytical method for determining bromate in drinking water was developed using liquid chromatography-tandem mass spectrometry (LC-MS/MS). The (18)O-enriched bromate was used as an internal standard. The limit of quantification (LOQ) of bromate was 0.2 μg/L. The peak of bromate was separated from those of coexisting ions (i.e., chloride, nitrate and sulfate). The relative and absolute recoveries of bromate in two drinking water samples and in a synthesized ion solution (100 mg/L chloride, 10 mg N/L nitrate, and 100 mg/L sulfate) were 99-105 and 94-105%, respectively. Bromate concentrations in 11 drinking water samples determined by LC-MS/MS were <0.2-2.3 μg/L. The results of the present study indicated that the proposed method was suitable for determining bromate concentrations in drinking water without sample pretreatment.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号