Effect of dopants on TL characteristics of LiF:Mg,Cu,P detectors |
| |
Authors: | ?. Kne?evi? M. Ranogajec-Komor S. Miljani? |
| |
Affiliation: | Ru?er Bo?kovi? Institute, Bijeni?ka 54, 10000 Zagreb, Croatia |
| |
Abstract: | LiF:Mg,Cu,P detectors produced at the Institute of Nuclear Physics in Krakow have shown very good dosimetric characteristics. Understanding of the effect of the concentration and type of dopants is important in the characterization of TL materials. The aim of work was to investigate the influence of the type and concentration of the dopants on the photon energy response of these detectors by irradiations “in air” and on the ISO water phantom in the range of mean photon energies between 33 and 164 keV. The influence of dopants on the glow curves, sensitivity and reproducibility was also examined. Results showed that measured energy dependence values are lower compared to the theoretical values both “in air” and on phantom. The type and concentration of the dopants influence the shape of the glow curves and sensitivity while for energy dependence is more important the presence of certain activators, namely copper. |
| |
Keywords: | |