首页 | 本学科首页   官方微博 | 高级检索  
     


Liquid chromatography tandem mass spectrometry method for determination of N-acetylcysteine in human plasma using an isotope-labeled internal standard
Authors:Lu Chuan  Liu Gangyi  Jia Jingying  Gui Yuzhou  Liu Yanmei  Zhang Mengqi  Liu Yun  Li Shuijun  Yu Chen
Affiliation:Central Laboratory, Shanghai Xuhui Central Hospital, 966 Middle Huaihai Road, Shanghai, China.
Abstract:A liquid chromatography–tandem mass spectrometry (LC‐MS/MS) method was developed to determine total N‐acetylcysteine in human plasma. Mass spectrometric detection was achieved in positive electrospray ionization and multiple reaction monitoring mode. The mass transition pairs of N‐acetylcysteine and the isotope‐labeled internal standard d3‐N‐acetylcysteine were 164 → 122 and 167 → 123, respectively. The method was linear over the range of 10–5000 ng/mL in human plasma. The adoption of trichloroacetic acid significantly enhanced the extraction recovery. The blank matrix was screened to minimize the influence of endogenous N‐acetylcysteine. After being fully validated, the method was successfully applied to the pharmacokinetic and bioequivalent study of N‐acetylcysteine after oral administration of 600 mg tablets to 24 healthy Chinese volunteers. Copyright © 2010 John Wiley & Sons, Ltd.
Keywords:N‐acetylcysteine  LC‐MS/MS  human plasma  bioequivalence
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号