首页 | 本学科首页   官方微博 | 高级检索  
     


A flow-injection system using immobilized peroxidase and chromogenic reagents for possible determination of hydrogen peroxide
Authors:Bo Olsson
Affiliation:(1) Department of Analytical Chemistry, University of Lund, Lund, Sweden
Abstract:Summary A flow injection system with a reactor containing peroxidase immobilized on porous glass was used for the determination of hydrogen peroxide. A number of chromogenic reagents were studied in combination with different immobilization methods, in particular with regard to adsorption of the coloured product on the enzyme support. A f. i. a. manifold had a throughout of 210 samples h–1 and a detection limit of 0.1 mgrM H2O2 and could be used for at least 17000 determinations with the same enzyme reactor. Systems with higher sensitivity or higher sample throughput are also described.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏