首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Quality assessment of liquid pharmaceutical preparations by HSS-GC-FID
Authors:A Mornar  M Sertić  B Nigović
Institution:1. Faculty of Pharmacy and Biochemistry, University of Zagreb, A. Kova?i?a 1, 10000, Zagreb, Croatia
Abstract:A new headspace gas chromatographic method with flame ionization detection (HSS-GC-FID) was developed and validated for the determination of methanol as the main volatile impurity present in ethanolic pharmaceutical preparations. The use of static headspace sampling minimized the interference of other volatile matrix components and provided satisfactory results in purity assessment of different complex samples. The developed procedure revealed to be rapid, sensitive, reproducible and accurate. The detection and quantification limits of methanol were 0.0003 and 0.0011% (v/v), respectively, and were sufficiently low to enable the estimation of organic volatile impurity according to the ICH guideline as well as the examination of methanol limit specified in European Pharmacopoeia for liquid pharmaceutical preparations. The proposed method was successfully applied to the analysis of diverse alcoholic herbal extracts and tinctures as well as ethanolic dermatological lotions.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏