Voltammetric behavior of rutin at a boron-doped diamond electrode. Its electroanalytical determination in a pharmaceutical formulation
Authors:P?nar Talay P?nar  Yavuz Yard?m  Zühre ?entürk
Institution:1292. Faculty of Science, Department of Analytical Chemistry, Yüzüncü Y?l University, 65080, Van, Turkey
2292. Faculty of Pharmacy, Department of Analytical Chemistry, Yüzüncü Y?l University, 65080, Van, Turkey
Abstract:This paper examined for the first time, the possibilities of the usage of a boron-doped diamond electrode for the redox behavior of rutin using cyclic and adsorptive stripping voltammetry. The cyclic voltammograms showed a pair of redox peaks at lower potentials followed by an irreversible oxidation peak at higher positive potential. Using square-wave adsorptive stripping voltammetry, the compound yielded a well-defined voltammetric response in Britton-Robinson buffer, pH 4.0 at +0.48 V (vs. Ag/AgCl) (after 60 s accumulations at a fixed potential of 0.2 V). The calibration curve was linear in the concentration range from 0.01 µg mL?1 to 0.1 µg mL?1 (1.64×10?8 M ? 1.64×10?7 M). A detection limit of 0.0017 µg mL?1 (2.78×10?9 M) was observed without any chemical modifications and electrochemical surface pretreatments. As an example, the practical applicability of boron-doped diamond electrode was tested with the measurement of rutin in dietary supplement products.  src=
Keywords:
本文献已被 SpringerLink 等数据库收录!