Effects of humidity and light on dosimetric properties of clear polymethylmethacrylate
Authors:H M Khan  G Ahmad  A Sattar
Institution:(1) National Centre of Excellence in Physical Chemistry, University of Peshawar, Pakistan;(2) Present address: Nuclear Institute for Food and Agriculture, Tarnab, Peshawar
Abstract:Effects of humidity and light conditions /fluoroscent light, sunlight and diffuse sunlight/ on the response of 2 mm thick clear polymethylmethacrylate /PMMA/ dosimeter during irradiation and during postirradiation storage were determined spectrophotometrically at 305 and 314 nm. The results showed that irradiation of dosimeters at 12–97% relative humidity did not result in any change in the response upto 2 weeks of post-irradiation storage. However, there is some decrease in the response at higher humidities />76%/ for longer storage time. Post-irradiation storage and dosimeters at constant relative humidity conditions between 35 and 97% had no effect on the response up-to 3 weeks. Post-irradiation storage at lower relative humidity /12%/ showed some decrease in the response for longer periods. The response of dosimeter is not very sensitive to short time exposure to diffuse sunlight and fluorescent light. However, a significant increase in the optical density was observed during exposure of dosimeters to direct sunlight.