Eine Mikromethode der Flammen-Atomabsorptions- und Emissionsspektrometrie (Platinschlaufen-Methode)—III: Verbesserung der analytischen Aussage durch zeitliche Auflösung der Impulssignale
Authors:H Berndt  J Messerschmidt
Institution:Institut für Spektrochemie und angewandte Spektroskopie, Bunsen-Kirchhoff-Str. 11, D-4600 Dortmund 1, Federal Republic of Germany
Abstract:By the application of storage oscillograph or a transient recorder in combination with a common chart recorder, the transient signals obtained with the loop method were time-resolved. This approach permitted a better analysis and interpretation of the phenomena, and allowed the statement of operating conditions that yield improved analytical performance including the power of detection. From the shape of the signal, conclusions could be drawn regarding the uniformity of the sample evaporation and the formation of compounds with differing evaporation behaviour.