首页 | 本学科首页   官方微博 | 高级检索  
     检索      


An optimized gas chromatographic determination of priority pollutant phenols
Authors:O H Masi  W M Gulick
Abstract:Priority pollutant phenols are determined by gas chromatography on an SE-30 WCOT column using hydrogen carrier and flame ionization detection with dual internal standards. Linear responses over ranges ca. 12–125 ng per component are achieved and detection limits at or below 1 ng are obtained. Carrier flow and temperature program are optimized to maintain baseline resolution while affording an analysis time of less than 15 minutes.
Keywords:Gas chromatography  Capillary columns  fused silica  Priority pollutant phenols  Internal standard quantitation