Comparison of pyrolysis-mass spectrometry with high performance liquid chromatography for the analysis of vancomycin in serum |
| |
Authors: | Ghassempour A Darbandi M K Asghari F S |
| |
Affiliation: | Center for Chemical Research, Faculty of Science, Shahid Beheshti University, Evin, Tehran, Iran. |
| |
Abstract: | Mass spectrometry coupled with a pyrolysis inlet system (Pyr-ms) is compared with high performance liquid chromatography (HPLC) for the determination of vancomycin and its crystal degradation products (CDP-Is) in human serum. Quantitative analysis of Pyr-ms was performed by selected ion monitoring (SIM) method at 108 mass:charge (m/z) of pyrolysis product of vancomycin. 3-Nitroaniline (138 m/z) was used as an internal standard. A mu-Bondapak C(18) column and the gradient mobile phase of triethylamine buffer (pH 6.2), acetonitrile and tetrahydrofuran and a photometric detection at 205 nm are found to be the optimum conditions for the HPLC determination of vancomycin and CDP-Is. The limit of detection (LOD=1 ng ml(-1)), linearity (1 ng ml(-1)-10 mg ml(-1)), relative standard deviation (R.S.D.=1%), time analysis (1/2 h) and sample volume (50 mul) of Pyr-ms are far better than of the HPLC method. However, the HPLC method can individually determine the concentration of vancomycin and its degradation products. |
| |