首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Potentiometric determination of dialysate urea nitrogen
Authors:Koncki R  Radomska A  G?ab S
Institution:Department of Chemistry, University of Warsaw, Pasteura 1, 02-093 Warsaw, Poland.
Abstract:An enzymatically modified ammonium ion-selective electrode has been applied for the determination of urea in spent dialysate. The biosensor has been used in a simple flow-injection analysis (FIA) system. The system enables one to perform over 25 dialysate urea nitrogen (DUN) determinations per hour. The interferences from other components of posthemodialysis fluid were eliminated by simultaneous measurements with non-modified enzymatically ion-selective electrode. It is possible to use both the sensors in a simplified differential potentiometric system. The results of DUN determination using the biosensor/FIA system and a conventional method of urea determination were comparable. The presented analytical system can potentially find wider biomedical application in the monitoring of hemodialysis progress.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏