Micromachined Enzyme Reactor for FIA System
Authors:Murakami Y  Suda M  Yokoyama K  Takeuchi T  Tamiya E  Karube I
Abstract:An enzyme reactor consisting of a 2.6-m-long silicon capillary with glucose oxidase immobilized on the inner surface was fabricated using micromachining techniques. A V-shaped groove of 100 μm width, formed by anisotropic etching, was anodically bonded to a glass plate to create the capillary. Glucose oxidase was covalently immobilized with 3-aminopropyltriethoxysilane and glutaraldehyde. The reactor was evaluated by connecting it to a Flow injection analysis system for glucose detection. Glucose concentrations were in the range of 10−3 to 5 × 10−2M with a volume of 0.2 μl of glucose solution.
Keywords: