Calibration and test of an aneroid mini-bomb combustion calorimeter
Institution:Centro de Investigação em Química, Department of Chemistry, Faculty of Science, University of Porto, Rua do Campo Alegre, 687, P-4169-007 Porto, Portugal
Abstract:A new mini-bomb combustion calorimeter designed at the University of Lund was improved, installed and calibrated at the University of Porto. This calorimeter is suitable for high precision combustion calorimetry with samples of mass about (10 to 40) mg. The energy equivalent of the calorimeter, εcal = (1946.45 ± 0.11) J · K−1, was obtained from 15 independent calibration experiments with benzoic acid SRM 39i. Anthracene, succinic acid, acetanilide, and 1,2,4-triazole were used as test compounds, with excellent agreement with the literature values.
Empty Cell-ΔcHmΔfHm(cr)
Empty CellkJ · mol−1kJ · mol−1
Anthracene7062.6 ± 2.1124.3 ± 2.8
Succinic acid1490.2 ± 0.7−941.3 ± 0.9
Acetanilide4226.2 ± 1.1−208.2 ± 1.6
1,2,4-Triazole1326.1 ± 0.4110.3 ± 0.5
Keywords:
本文献已被 ScienceDirect 等数据库收录!