首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Liquid chromatography tandem mass spectrometry method for the quantification of amisulpride with LLOQ of 100 pg/mL using 100 microL of plasma
Authors:Nirogi Ramakrishna  Bhyrapuneni Gopinadh  Kandikere Vishwottam  Mudigonda Koteshwara  Ajjala Devender  Suraneni Ramakrishna  Mukkanti K
Institution:Pharmacokinetics and Drug Metabolism, Drug Discovery, Suven Life Sciences Ltd, Serene Chambers, Road 5, Avenue 7, Banjara Hills, Hyderabad 500034, India. ramakrishna_nirogi@yahoo.co.in
Abstract:A sensitive and selective high-performance liquid chromatography-positive ion electrospray tandem mass spectrometry method was developed and validated for the quantification of amisulpride in 100 microL of human plasma. Following liquid-liquid extraction, the analytes were separated using an isocratic mobile phase on a reverse-phase column and analyzed by MS/MS in the multiple reaction monitoring mode using the respective (M + H)(+) ions, m/z 370-242 for amisulpride and m/z 341-112 for the internal standard. The assay exhibited a linear dynamic range with a lower range of 0.1-100 ng/mL and a higher range of 1-500 ng/mL of amisulpride in human plasma. The lower limit of quantification was 0.1 ng/mL with a relative standard deviation of less than 10%. Acceptable precision and accuracy were obtained for both linearity ranges. A run time of 2.0 min for each sample made it possible to analyze more than 275 human plasma samples per day. The validated method has been successfully used to analyze plasma samples for application in pharmacokinetic studies.
Keywords:amisulpride  liquid chromatography–tandem mass spectrometry  plasma
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号