Flow injection on-line ultrasound-assisted extraction of iron in meat samples coupled to a flame atomic absorption spectrometric system
Authors:A.?Moreno-Cid,M.?C.?Yebra  author-information"  >  author-information__contact u-icon-before"  >  mailto:qncayebi@usc.es"   title="  qncayebi@usc.es"   itemprop="  email"   data-track="  click"   data-track-action="  Email author"   data-track-label="  "  >Email author,S.?Cancela,R.?M.?Cespón
Affiliation:(1) Department of Analytical Chemistry, Nutrition and Bromatology, Faculty of Chemistry, University of Santiago de Compostela, 15782 Santiago de Compostela, Spain
Abstract:Iron was extracted on-line from solid meat samples by a simple and rapid continuous ultrasound-assisted extraction system (CUES). The CUES is connected to a flow injection manifold, which allows the on-line flame atomic absorption spectrometric determination of iron. A Plakett–Burman design was used for the optimisation of the CUES. The method achieved a total sampling frequency of 11 samples per hour with a relative standard deviation for the complete procedure of 0.4%. The detection limit was 0.6 mgrg g–1 (dry mass) for a sample amount of 30 mg. Accurate results were obtained by measuring the certified reference materials BCR-186 (pig kidney) and BCR-184 (bovine muscle). The analytical procedure was applied to different real meat samples with satisfactory results.
Keywords:Continuous ultrasound-assisted extraction system  Flame atomic absorption spectrometry  Iron determination  Meat samples