首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of polynuclear aromatic hydrocarbons in water samples using large volume on-column injection capillary gas chromatography
Authors:Paul L Morabito  Terry McCabe  Joseph F Hiller  Don Zakett
Abstract:An automated large volume on-column injection technique for capillary gas chromatography (GC) with solvent divert and heated retention gap technology has been utilized to determine polynuclear aromatic hydrocarbons (PAHs) in samples of industrial plant process water. Injecting large sample volumes on-column enabled the sample preparation procedure to be simplified and provided a fast, labor-saving technique for screening water samples. Diverting approximately 95% of the solvent away from the analytical column and the detector enabled chromatography to reflect classical capillary loading and detector conditions. Simplifications include significant reduction of sample and eluent volumes used during extraction and the elimination of Kuderna-Danish evaporative concentration. System performance, such as linearity and limit of detection, were evaluated for selected PAHs. Spiked water samples were prepared in the lower μ/L range to determine extraction efficiency. Results are compared with those obtained by a reputable contract laboratory following EPA Method 625.
Keywords:On-column injection  Large (100 μ  l) injections  Heated retention gap  Sample preparation  Extraction efficiency  Linearity  Limit of detection  Polynuclear aromatic hydrocarbons (PAHs)  Industrial Waste  Water analysis
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号