A comparative test of differential scanning calorimeters
Authors:P. J. van Ekeren  C. M. Holl  A. J. Witteveen
Affiliation:(1) Dutch Society for Thermal Analysis TAWN, Utrecht University, Padualaan 8, NL-3584 CH Utrecht;(2) Thermodynamic Center, Dept. of Interfaces and Thermodynamics, Utrecht University, Padualaan 8, NL-3584 CH Utrecht;(3) Dept. RGL, Akzo Nobel Central Research, P. O. Box 9300, NL-6800 SB Arnhem, The Netherlands
Abstract:The Dutch Society for Thermal Analysis has developed tests to measure the resolution and the sensitivity of Differential Scanning Calorimeters. For this test the substance 4,4prime-azoxyanisole is used. This substance shows two transitions: a solid to liquid crystal transition at about 117°C (DeltaHap120 J g–1) and a liquid crystal to isotropic liquid transition at about 134°C (DeltaHap2 J g–1). The resolution test is performed using an amount of 5 mg substance and a high heating rate of 20°C min–1. The resolution is evaluated by measuring how well the two peaks are separated. An amount of 0.25 mg substance and a low heating rate of 0.1 °C min–1 is used for the sensitivity test. The sensitivity is evaluated as the ratio of the peak height of the LC-transition and the top-top noise level.Members of the TAWN were asked to participate in the test. Each participant was provided with an amount of sample and a test procedure. 47 Contributions were received and these results are presented.All members of the TAWN who participated in the tests are thanked for their contributions.
Keywords:DSC  resolution  sensitivity  test