首页 | 本学科首页   官方微博 | 高级检索  
     


Modelling the gas chromatographic separation of multicomponent samples on a system of three open-tubular columns coupled in series
Authors:T. Hevesi, J. Krupí  k  J. R. Chré  tien
Affiliation:(1) Department of Analytical Chemistry, Faculty of Chemical Technology, Slovak Technical University, Radlinského 9, SK-812 37 Bratislava, Slovakia;(2) Laboratoire de Chimie Analytique Industrielle et Chimiométrie, Faculté des Sciences, Université d'Orleáns, B.P. 6759, F-45067 Orleáns Ce"cacute"ex 2, France
Abstract:A procedure was developed for modelling the gas chromatographic separation on a system of columns coupled in series. This procedure can be used for computer based optimization of lengths and order of serially coupled columns at isothermal conditions. The sample component retention factors and column resistances needed for the model can easily be measured on each individual column. The proposed procedure was verified by analyzing a 52 component hydrocarbon mixture on three columns of different polarity coupled in series by various column orders. Good agreement between experimental and calculated retention data was found. The procedure is also well suited for optimization of the chromatographic selectivity by coupling columns with different selectivity in series.Dedicated to Professor Dr. h.c. mult. J.F.K. Huber on the occasion of his 70th birthday
Keywords:
本文献已被 SpringerLink 等数据库收录!