Standardisation and Quality Assurance |
| |
Authors: | M. J. Richardson |
| |
Affiliation: | (1) Polymer Research Centre, University of Surrey, Guilford Surrey, GU2 5XH, UK |
| |
Abstract: | Modern DSC instruments generally give very reproducible results that are ideal for quality assurance work. However, various corrections are needed before quantitative data are obtained. Since the corrections differ from one type of instrument to another, it is important to ensure that they are all properly carried out. Without these corrections, it is fruitless to compare data from different sources. The types of problem, and their solutions, are discussed. This revised version was published online in August 2006 with corrections to the Cover Date. |
| |
Keywords: | calibrants quality assurance quantitative data standard procedures |