Artifacts in T1 rho-weighted imaging: compensation for B(1) and B(0) field imperfections
Authors:Witschey Walter R T  Borthakur Arijitt  Elliott Mark A  Mellon Eric  Niyogi Sampreet  Wallman Daniel J  Wang Chenyang  Reddy Ravinder
Institution:Graduate Group in Biochemistry and Molecular Biophysics and MMRRCC, University of Pennsylvania, B1 Stellar-Chance Laboratories, 422 Curie Boulevard, Philadelphia, PA 19104-6100, USA. witschey@med.upenn.edu
Abstract:The origin of spin locking image artifacts in the presence of B(0) and B(1) magnetic field imperfections is shown theoretically using the Bloch equations and experimentally at low (omega(1) < Delta omega(0)), intermediate (omega(1) approximately Delta omega(0)) and high (omega(1) > Delta omega(0)) spin locking field strengths. At low spin locking fields, the magnetization is shown to oscillate about an effective field in the rotating frame causing signature banding artifacts in the image. At high spin lock fields, the effect of the resonance offset Deltao mega(0) is quenched, but imperfections in the flip angle cause oscillations about the omega(1) field. A new pulse sequence is presented that consists of an integrated spin echo and spin lock experiment followed by magnetization storage along the -z-axis. It is shown that this sequence almost entirely eliminates banding artifacts from both types of field inhomogeneities at all spin locking field strengths. The sequence was used to obtain artifact free images of agarose in inhomogeneous B(0) and B(1) fields, off-resonance spins in fat and in vivo human brain images at 3 T. The new pulse sequence can be used to probe very low frequency (0-400 Hz) dynamic and static interactions in tissues without contaminating B(0) and B(1) field artifacts.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号