Simultaneous biphasic liquid scintillation determination of americium and plutonium |
| |
Authors: | A M Bomben H E Equillor A A Oliveira |
| |
Institution: | (1) Centro Atómico Ezeiza-CC, Gerencia de Seguridad Radiológica y Nuclear, Comisión Nacional de Energía Atómica, 40-Aeroquerto de Ezeiza, República Argentina |
| |
Abstract: | A procedure for the simultaneous determination of241Am and239Pu or238Pu was carried out in samples such as air filters, sweep-tests, aqueous solutions and urine. The method described here includes a previous treatment of the samples in accordance with the type of matrix in which the actinides are included and a liquid scintillation counting using a two-phase cocktail. The upper detection limit was estimated to be 50 mBq for a 50-minute counting time. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |