The Electropolymerization of 2,6‐Diaminopyridine and Its Application as Mercury Ion Selective Electrode |
| |
Authors: | Fariba Bakhtiarzadeh,Sulaiman Ab Ghani |
| |
Abstract: | A composite graphite (CG) electrode modified with poly(2,6‐diaminopyridine) (PDAP) was used as solid state‐ion selective electrode for determination of mercury. The electrooxidation of monomer 2, 6 diaminopyridine (DAP) onto CG was accomplished from the 30 mM DAP in 5% H2SO4 and 0.5 M ZnSO4. The electrode displayed Nernstian response with slope of 28.4±1 mV decade−1 in concentration range of 1×10−6 to 1×10−1 M and in solution of pH 3–5. The limit of detection for electrode was 3×10−8 M with response time of 25 s. The electrode was also suitable as an indicator electrode in the potentiometric titration of Hg2+ with iodide. |
| |
Keywords: | Electropolymerization 2,6‐Diaminopyridine Mercury Solid‐state ion selective electrode |
|
|