6.
Summary An analytical procedure for the determination of Se, as well as of other trace elements of clinical interest, in undiluted blood serum samples is presented. This procedure allows the on-line wet ashing of the sample, the hydride generation and the determination of the SeH
2 formed in one step. For this purpose, carrier solution and the injected sample (300 l) were merged with an acid stream (80% H
2SO
4, 12% HClO
4, 8% HNO
3) and passed through a reaction coil heated up to 240°C. To increase the dispersion, the wet digestion was carried out in an ultrasonic field. After trapping of the gas bubbles, 5 mol/l HCl and 0.053 mol/l NaBH
4 solutions were added via T-junctions for Se-hydride generation. The nebulizing system of the ICP-OES was used as gas/liquid separator. The hydride was swept from the spray chamber into the plasma by an argon gas flow with droplets containing other sample constituents. This allows the simultaneous determination of other trace elements of clinical interest, for example Fe, Cu and Zn. The relative detection limit for Se in blood serum was 5 g/l. The quality of the developed procedure was verified in two ways: by measuring SRM Seronorm 116 and by comparing the results for different serum samples with the results obtained with ETAAS. Our results were in good agreement in both cases.Dedicated to Professor Dr. Wilhelm Fresenius on the occasion of his 80th birthday
相似文献