首页 | 本学科首页   官方微博 | 高级检索  
相似文献
 共查询到20条相似文献,搜索用时 31 毫秒
1.
An experimental calibration of two common nuclear track detectors, CR-39 and Lexan, for gamma and electrons radiation, was performed using various irradiation facilities. The dose response was obtained as a function of two parameters, the bulk etch rate and the UV absorbance for a wide dose range from 10 to 1000 kGy. The bulk etch rate sensitivity, for gamma and electrons, in CR-39 detector is higher than for Lexan detector. Lexan has a well-defined UV absorbance spectrum, but presents saturation for doses higher than 500 kGy, the same saturation characteristic is observed for the corresponding bulk etch rate response. For electron and gamma radiation, CR-39 shows a good response for doses from 10 kGy up to 1000 kGy, where data fit well an exponential curve for electrons and a lineal curve for gamma radiation.  相似文献   

2.
The effects of Co-60 gamma-rays and about 25-MeV pulsed electrons have been investigated on CR-39 plastic nuclear track detector. The bulk etch rate was measured as a function of the absorbed dose under three different irradiation conditions; gamma-ray exposure in air at the dose rate of 6 kGy/h, that in vacuum and electron shower in air at 126 kGy/h on average. It was found that the bulk etch rate (V) increased nearly exponentially with the dose (D); V=V0exp(kD), and the characteristic parameter (k) was 35, 9 and 12 MGy−1, respectively. It is considered that a lack of oxygen in damage formation processes could occur momentarily during a short electron pulse, just similarly to the vacuum effect of reduction in track etch rates in outgassed CR-39 plastic.  相似文献   

3.
It is well established that the bulk etch rates for solid state nuclear track detectors are affected by the concentration and the temperature of the etchant. Recently, we found that the bulk etch rate for the LR 115 detector to be affected by stirring during etching. In the present work, the effects of stirring on the bulk etch rate of the CR-39 detector is investigated. One set of sample was etched under continuous stirring by a magnetic stirrer at 70°C in a 6.25 N NaOH solution, while the other set of samples was etched without the magnetic stirrer. After etching, the bulk etch thickness was measured using Form Talysurf PGI (Taylor Hobson, Leicester, England). It was found that magnetic stirring did not affect the bulk etch of the CR-39 detector, which was in contrast to the results for the LR 115 detector.  相似文献   

4.
The effect of gamma-rays and pulsed electrons has been investigated on the bulk etch rate of CR-39 detector at doses up to 100 kGy under various dose-rate between 0.0044 and 35.0 Gy/s. The bulk etch rate increased exponentially with the dose at every examined dose-rates. It was reveled to be strongly depend on the dose-rate: the bulk etch rate was decreased with increasing dose-rate at the same total dose. A primitive model was proposed to explain the dose-rate effect in which oxygen dissolved was assumed to dominate the damage formation process.  相似文献   

5.
The effect of 5-15 keV electron irradiation on the etch rate of CR-39 nuclear track detectors has been investigated for surface doses from 0.09 to 13.85 J/cm2. Even before etching, electrons produce a surface depression proportional to the surface dose raised to the power 0.78 +/- 0.01. Etch rate enhancement was observed in the surface region (about 4 micrometers deep) and was well approximated by a Gaussian function of depth. The depth and thickness of the etch rate enhancement increase with increasing electron energy, but are independent of dose. The maximum etch rate is proportional to the peak volumetric dose raised to the power of 1.45 +/- 0.06, independent of electron energy. No signs of a saturated response were found in the etch rate. This enhanced etching rate may significantly affect nuclear track identification in situations with an electron background.  相似文献   

6.
The effect of infrared radiation on the etching characteristics of CR-39 plastic track recorder has been studied. CR-39, which is commonly employed in cosmic rays studies is found to be affected by IR radiations. The changes in the bulk etch rate, in the track etch rate and in etching efficiency due to infrared radiation exposure of CR-39 are discussed on the basis of scission and cross linking during infrared exposure.  相似文献   

7.
Track etch rate characteristics of CR-39 plastic detector exposed to 28Si ions of 670 MeV energy have been investigated. Experimental results were obtained in terms of frequency distribution of the track diameter, track density and bulk etching rate. A dependence of the mean track diameter on energy was found. The application of the radiation effect of heavy ions on CR-39 in the field of radiation detection and dosimetry are discussed. Results indicated that it is possible to produce etchable tracks of 28Si in this energy range in CR-39. We also report the etching characteristics of these tracks in the CR-39 detector.  相似文献   

8.
The effect of additives and cure cyle parameters on pre and post irradiation annealing characteristics of CR-39 detectors has been investigated. Effect of pre and post irradiation annealing on bulk etch rate, sensitivity, track diameter and etching behaviour of various CR-39 detectors has been investigated.  相似文献   

9.
In this work, we have studied the effect of the radiation damage caused by the incident particles on the activation energy of etching for CR-39 samples. The damage produced by the incident particle is expressed in terms of the linear energy transfer (LET). CR-39 samples from American Acrylic were irradiated to three different LET particles. These are N (LET200 = 20 KeV/μm) as a light particle, Fe (LET200 = 110 KeV/μm) as a medium particle and fission fragments (ff) from a 252Cf source as heavy particles. In general the bulk etch rate was calculated using the weight difference method and the track etch rate was determined using the track geometry at various temperatures (50–90 °C) and concentrations (4–9 N) of the NaOH etchant. The average activation energy Fb related to the bulk etch rate vb was calculated from ln vb vs. l/T. The average activation energy Et related to the track etch rate vt was estimated from ln vt vs. l/T. It is shown that activation energy of etching is a constant value for CR-39 detector where Et was found to be independent on the damage produced by the incident particle.  相似文献   

10.
Abstract

The effects of neutron, gamma and alpha radiations on the alpha and fission fragment tracks registration and revelation properties of CR-39 detectors (CR-39 and CR-39(DOP) were studied. It was found that the ratio of the bulk etch rate of irradiated to unirradiated (VG(irr.)/VG(unirr.) detectors is linearly dependent on dose. An exponential decrease in fission track densities with increase in neutron fluence was observed. The ratio of VG(irr.)/VG(unirr.) was found to be high in CR-39 than that in CR-39(DOP) exposed to the same reactor neutron fluence. The decrease in fission track densities with increase in neutron fluence was observed to be faster in CR-39 than in CR-39(DOP). This indicates that doping with dioctyl phthalate improves the radiation resistance of CR-39 detectors. It was observed that in detectors exposed to an alpha flux of the order of 9.36 × 106 / cm2, the fission track density was reduced by 11% and thereafter it remained constant. The results also indicate that thermal neutron fluence up to 7.01 ×1011 neutrons/cm2 does not affect the alpha and fission track densities. I.R. spectra were also studied to find out the nature of chemical changes produced by these radiations on CR-39.  相似文献   

11.
Solid state nuclear track detectors are commonly used for measurements of concentrations of radon gas and/or radon progeny. All these measurements depend critically on the thickness of the removed layer during etching. However, the thickness of removed layer calculated using the etching period does not necessarily provide a sufficiently accurate measure of the thickness. For example, the bulk etch rate depends on the strength of stirring during etching for the LR 115 detector. We propose here to measure the thickness of the removed layer by using energy-dispersive X-ray fluorescence spectrometry. In the present work, a reference silver nitrate pellet is placed beneath the LR 115 detector, and the fluorescence X-ray intensity for silver is then measured. We have found a linear relationship between the X-ray intensity and the thickness of the removed layer for LR 115 detector. This provides a fast method to measure the thickness of removed layer from etching of LR 115 detector. However, this method was found to be inapplicable for the CR-39 detector. Therefore, alternative methods have yet to be explored for the CR-39 detector.  相似文献   

12.
The effect of post exposure annealing on the track registeration characteristics of CR-39 (Pershore) Solid State Nuclear Track Detectors has been studied. The changes introduced in the bulk etch rates, track diameter, and the track density have been observed. The results indicate that the bulk etch rate of the detectors and the size of the tracks are considerably changed due to annealing. Variations in the track density are, however, not serious in the presently studied temperature range.  相似文献   

13.
A review was given for our recent studies on the latent tracks in CR-39 nuclear track detector. The radial size of track core has been determined through UV spectral measurements combined to the model of track overlapping and by AFM observations of slightly etched detectors. The track core radius was found to be about a few nano-meters and almost proportional to the cubic root of stopping power. As a control study, the etching properties of CR-39, irradiated by low-LET radiation, has been examined. The observed depth dependence and dose-rate dependence of the bulk etch rate of the irradiated CR-39 were explained that the damage formation process was governed by the reaction between the radiation induced radicals and the oxygen supplied from the air. This indicated that latent tracks in CR-39 are produced through local radiation induced oxidation process along the ion paths. Studies on vibration spectra, near-IR, FT-IR and Raman spectra, of CR-39 have also been carried out to estimate the chemical structure of the latent tracks. The creation of OH group in irradiated CR-39 has been confirmed.  相似文献   

14.
Computation of the neutron response of CR-39 detectors needs to simulate the track formation by neutron induced charged particles taking into account the bulk etch rate and the track etch rate varying along the particle trajectories. The latter one was determined experimentally by track length measurement. The results allowed to derive the relationship between the track etch rate and the restricted energy loss of the charged particles. On this basis, the geometrical track parameters and track etch rates as well as the critical angle of particle incidence could be determined for protons and alpha particles in the energy range from 0.2 to 8.8 MeV. The energy dependence of the critical angle enabled to determine the detection efficiency for a charged particle of given energy and direction.  相似文献   

15.
《Radiation measurements》1997,28(1-6):455-462
Using CR-39 plastic track detectors the range values of 16O ions at two different energies (initially in the beam line, 39.97 MeV/n and 69.98 MeV/n) were measured after escaping the beam pipe and found to be (3050 ± 40) μm and (8210 ± 90) μm, respectively. The longitudinal and projected angular spread of oxygen ions of an initial energy of 69.98 MeV/n in the region of the Bragg peak was derived from the measured geometrical parameters of tracks. Based on a calibration curve (etch rate ratio vs total linear energy transfer in CR-39) and the measured track length distribution at the range end of oxygen ions, the complete depth dose profile of a 67.7 MeV/n 16O beam in CR-39 (plateau, extended Bragg peak and residual ionization caused by projectile like fragments) was obtained.  相似文献   

16.
A new method based on the use of the confocal microscope is described in order to evaluate the CR-39 response function for Li-7 ions with an incident energy of 10.77 MeV. This method uses the formulations developed by Fromm et al. and considers two etching velocities: VB represents the bulk etch rate and remains constant, and VT the track etch rate, which varies along the particle's path. The confocal microscope seems to bring big improvements for track analysis. The first results of VT versus the particle range are presented and compared with the curves obtained by the sequential etching method. The obtained VT are plotted and compared to LET, REL350 and the cumulative radial dose.  相似文献   

17.
The bulk etch rate for two types of CR-39 detector was measured as a function of temperature and the activation energies of bulk etching was determined. Experimental values of track etch rate were derived directly from the function of the succesive measured track length vrs. etching time for 209Bi, 129Xe and 20Ne ions.

The maximum etchable length of 13 MeV/u 209Bi and 13.04 MeV/u 129Xe ions have been measured at and below these energies. A comparison of the measured and calculated track length data is presented.  相似文献   


18.
《Radiation measurements》2009,44(2):173-175
This work presents a novel method for determining bulk etch rate of CR-39 during prolonged etching by masking the surface with a ferrofluidic film held in position by magnetostatic forces. The CR-39 etching conditions were 6.25 M NaOH solution for 24 h at temperatures ranging from 50 to 80 °C. After etching, the heights of the resulting un-etched plateaus were measured using a Talyscan 150 profilometer. The removed layer thicknesses ranged from 12 to 85 μm, giving corresponding bulk etch rates in the range 0.5–3.54 μm/h.  相似文献   

19.
We studied the track response of CR-39 plastic nuclear track detectors (PNTD) for low (<6 MeV/n) and high (>100 MeV/n) energy heavy ions using the atomic force microscope (AFM). CR-39 PNTD was exposed to several heavy ion beams of different energy at HIMAC (Heavy Ion Medical Accelerator in Chiba). For AFM measurement, the amount of bulk etch was controlled to be ∼2 μm in order to avoid etching away of short range tracks. The response data obtained by AFM for ∼2 μm bulk etch was in good agreement with data obtained by the conventional optical microscope analysis for larger bulk etch. The response data from low energy beams (stopping near the surface) was also consistent with the data from high energy beams (penetrating the detector) as a function of REL (restricted energy loss) with the δ-ray cut off energy of ω0 = 200 eV. We experimentally verified that REL (ω0 = 200 eV) gives a universal function for wide energy range in CR-39 PNTD. This work has been done as part of a basic study in the measurement of secondary short range tracks produced by target fragmentation reactions in proton cancer therapy fields.  相似文献   

20.
《Radiation measurements》2007,42(9):1507-1509
Systematic studies have been made concerning the effects of gamma dose on the track registration properties of CR-39 plastics solid-state track detector (SSTD). The changes in bulk and track etch rates as well as in sensitivity due to gamma rays in the dose range of 0103kGy at different temperatures (338–353 K) are discussed on the basis of degradation and cross-linking mechanisms. The increase in bulk and track etch rates with gamma dose has been found while a rapid fall off in sensitivity at higher dose is also seen. From these measured values of bulk and track etch rates, the activation energies associated with them have also been calculated which show a decrease from unirradiated to 103kGy gamma dose.  相似文献   