 共查询到19条相似文献,搜索用时 78 毫秒
1.
设计了一种用于毛细管电泳系统的流动注射-负压进样装置。样品由蠕动泵输送到进样阀后再由缓冲液带到分离毛细管入口,由毛细管出口端施加的负压引入。进样时间由自制精密控时电路控制,经进样条件的优化,能获得良好的重现性。实验中两种阳离子峰面积和迁移时间的RSD(n=8)≤2.7%,优于传统重力进样,而且操作简便;与非接触电导检测器组装成流动注射-毛细管电泳系统,可实现快速、高效的在线分析。初步应用于无机阳离子的分离,取得了满意的结果。  相似文献   

2.
高效毛细管电泳的扩散进样   总被引:3,自引:0,他引:3  
陈义  竺安 《色谱》1991,9(6):353-356
]本文提出了用扩散作为高效毛细管电泳的进样方法,解释了此法的原理。用扁管区带电泳法研究了扩散进样对电泳效率的影响、歧视问题及其抗样品基质干扰的能力,并与常用的电迁移法和虹吸法进行了比较,指出了扩散进样的缺点。  相似文献   

3.
在毛细管电泳的胶束电动色谱(MEKC)模式下,采用压力辅助电动进样(PAEKI)的进样方式在线富集4种酚类雌激素(PEs)。对影响PAEKI的进样电压、进样时间等进行考察,并与传统的压力进样比较。结果表明,在最优的PAEKI条件下(-9 kV,0.3 psi(约2.1 kPa),0.4 min),4种PEs在7 min内基线分离,线性关系良好,相关系数(r)大于0.9936,己烷雌酚和双烯雌酚的线性范围为0.05~5 mg/L、双酚A和己烯雌酚的线性范围为0.1~10 mg/L;检出限(S/N=3)为0.0071~0.017 mg/L,富集倍数为11~15。使用该MEKC-PAEKI法对自来水和湖水水样进行测定,得到定量限(S/N=10)分别为0.029~0.064 mg/L和0.033~0.079 mg/L;加标回收率为75.6%~110.1%,相对标准偏差(n=5)为4.6%~11.8%。PAEKI不需要使用其他试剂,只需对电泳仪的参数进行适当调整即可实现对分析物的在线富集,简单、快速、自动化程度高。  相似文献   

4.
研究了用毛细管区带电泳法快速测定复方布洛芬片中布洛芬和伪麻黄碱含量的方法。在0.025mol/L的磷酸盐缓冲液(pH8.1)中,上述两组分可在3min内得以完全分离,用紫外检测器在210nm处检测,并以外标法定量。11次测定含有9.5mg/L盐酸伪麻黄碱和66.7mg/L布洛芬的试样溶液,相对标准偏差为2.9%(伪麻黄碱)和1.9%(布洛芬),回收率为103.1%(伪麻黄碱)和97.6%(布洛芬)。应用毛细管区带电泳法测定复方布洛芬片剂的含量,所得结果与HPLC法一致。  相似文献   

5.
毛细管区带电泳法快速测定复方布洛芬片的有效成分   总被引:3,自引:0,他引:3  
陈恒武  黄棣华  陈青俊  李会林 《色谱》1998,16(4):289-292
 研究了用毛细管区带电泳法快速测定复方布洛芬片中布洛芬和伪麻黄碱含量的方法。在0.025mol/L的磷酸盐缓冲液(pH8.1)中,上述两组分可在3min内得以完全分离,用紫外检测器在210nm处检测,并以外标法定量。11次测定含有9.5mg/L盐酸伪麻黄碱和66.7mg/L布洛芬的试样溶液,相对标准偏差为2.9%(伪麻黄碱)和1.9%(布洛芬),回收率为103.1%(伪麻黄碱)和97.6%(布洛芬)。应用毛细管区带电泳法测定复方布洛芬片剂的含量,所得结果与HPLC法一致。  相似文献   

6.
流动注射-毛细管电泳联用及应用进展   总被引:1,自引:0,他引:1  
流动注射是一种高效进样及在线溶液处理手段。毛细管电泳是一种高分离效率、高选择性的分析技术,但传统的毛细管电泳间歇式进样方式效率低且难用于过程分析,将流动注射进样技术与毛细管电泳结合,既弥补了毛细管电泳的进样缺陷,又可兼具两者的优点。有关两种技术的联用一直都在探索之中。文中对近年流动注射一毛细管电泳联用及应用研究进行了综述。  相似文献   

7.
毛细管区带电泳中场增强进样柱内富集的非线性特征   总被引:1,自引:0,他引:1  
直接柱头场效应进样是一种毛细管区带电泳柱内富集,其进样过程中样品在柱内的分布可分为两部分,即在运行缓冲溶液中的堆积区段和由电渗流引入的样品溶液区段.通过对溶质输运行为的研究表明:两区段长度与进样时间之间并非简单的线性关系,因此进样量与进样时间的关系也非线性,且与溶质淌度有关;进样量的增加并不能导致富集倍数的同步增加,由于层流的作用使得场效应进样柱内富集效果降低.为了在保持柱效基本不变情况下得到好的富集效果,除需使溶质在运行缓冲溶液和样品溶液中的电导率比极大外,进样时间也应与之匹配.  相似文献   

8.
流动注射与毛细管电泳联用在中药厚朴分析中的应用研究   总被引:7,自引:2,他引:5  
对流动注射与毛细管电泳联用(FI-HPCE)技术在中药厚朴分析中的应用进行了研究。采用了一种新的联用接口,并运用单纯形法优化了缓冲液浓度、pH值及电压、甲醇含量以取得最佳分离效果,使中药厚朴中的有效成分厚朴酚及和厚朴酚与其他杂质得以基线分离。采用优化的条件对中药厚朴及其部分成药进行了分离分析。采样频率10个样/h,精密度为2.1%,回收率为96% ̄102.9%。  相似文献   

9.
研究了流动注射在线固相萃取与毛细管区带电泳联用系统(FI-SPE-CZE)应用于测定血浆中痕量盐酸伪麻黄碱的方法。通过脱线和在线试样净化、优化CZE分离条件等措施消除了血浆中共存组份的干扰;通过增加固相萃取的进样量、加大场放大电动进样量和柱上堆积的力度等,提高了联用系统的灵敏度。血浆试样沉淀蛋白后可连续进入FI-SPE-CZE系统进行自动的浓缩、分离与分析,检测限为12ng/mL,分析速度达10次/h.应用本方法成功地测定了志愿者口服正常剂量盐酸伪麻黄碱后的血药浓度,并试用于药物动力学的研究。  相似文献   

10.
毛细管电泳 (CE)正向着实际体系分析和微型化分析的方向发展[1,2 ] .为提高分离效率 ,CE的进样量一般很小 (1 0 - 12 ~ 1 0 - 9L ) ,因此常采用在线样品富集方式以提高检测灵敏度 .电场放大进样 (FASI)是 CE中实现在线样品富集的简便方式[3] ,随着 CE和检测技术的发展 ,其应用范围和研究体系不断扩大[4 ] .三联吡啶钌 [Ru(bpy) 2 + 3]电化学发光 (ECL )作为一种高灵敏、高选择性的检测方法被逐渐应用于 CE分析 (CE-ECL) .我们对这种技术进行了研究 ,并将其应用于临床分析[5~ 7] .实验中发现 ,在施加正高压的条件下 ,不同进样条…  相似文献   

11.
The characteristics of bias caused by split-flow electrokinetic injection (SEKI), a new type of sample injection method used in coupled flow injection-capillary electrophoresis system (FI-CE), was investigated using pseudoephedrine hydrochloride, a basic drug, and ibuprofen, an acidic drug, as model analytes. It was found that bias imposed by SEKI under the condition of continuous sample matrix/running buffer was similar to that done by electrokinetic injection (EKI). The linearity of calibration curve provided by SEKI was similar to that offered by non-bias hydrodynamic injection (HDI) but significantly better than that obtained by EKI. These features were exploited to improve analytical performances in simultaneous determination of the minor ingredient of pseudoephedrine hydrochloride and the major ingredient of ibuprofen in a pharmaceutical preparation. Detectability of 0.7 mg/l for pseudoephedrine hydrochloride was achieved at a sample throughput rate of 24 times per hour, which is 30% lower than that obtained by HDI-based conventional CE. Relative standard deviations (RSDs) of 2.8% for the minor ingredient and 1.2% for the major ingredient were produced in 11 runs of a test solution containing 13.1 mg/l pseudoephedrine hydrochloride and 81.4 mg/l ibuprofen. This is an improvement compared to that obtained by HDI-based conventional CE. Analytical results for two batches of compound ibuprofen tablets by the SEKI-based FI-CE approach were in good agreement with that obtained by a conventional high performance liquid chromatographic method. __________ Translated from Chinese Journal of Chromatography, 2005, 23(2) (in Chinese)  相似文献   

12.
姚巍  徐淑坤 《分析化学》2002,30(7):836-838
采用流动注射与毛细管电泳联用(FI-CE)接口进行自动连续采样。对检测波长、电渗流改性剂浓度、载流中四硼酸钠浓度、载流PH、载流流速及采样环体积等条件进行优化,并对环境水样中常见阴离子的干扰进行研究,建立了同时测定水中硝酸根和亚硝酸根的方法。在采样量为50μl时硝酸根和亚硝酸根的检出限(3σ)分别为0.2和0.4mg/L;峰高RSD分别为1.6%和2.0%(20mg/L,n=25),峰面积RSD分别为1.7%和1.6%(20mg/L,n=25),采样频率为60/h。用本法测定井水水样的结果与离子色谱法一致。水样中加入10mg/L硝酸根和亚硝酸根的回收率为98%-108%。  相似文献   

13.
建立了基于水塞联用场放大进样(FESI)的区带毛细管电泳(CZE)检测多种样品中三聚氰胺的分析方法。水塞组成为40%乙腈和60%水,水塞进入时间200 s,进水压力3 kPa。以120 mmol/L NaH2PO4缓冲液(pH 2.2)-10%甲醇为运行缓冲溶液,以0.10 mmol/L NaH2PO4(pH 2.2)-20%乙腈为样品基体溶液,进样电压20 kV,进样时间80 s,分离电压20 kV。在优化实验条件下,与普通的CZE法比较,三聚氰胺的紫外检测灵敏度提高了800倍,检出限(S/N=3)由2.0 mg/L降至2.5μg/L,线性范围为10~1 000μg/L。将该方法用于多种样品中三聚氰胺残留的检测,回收率为98%~106%,相对标准偏差(RSD,n=4)均不高于5.1%。该方法克服了紫外检测灵敏度低的缺陷,具有检测灵敏、简便易行、预处理简单、干扰少、经济环保和适用范围广等优点。  相似文献   

14.
低电渗流毛细管区带电泳分离芳香胺   总被引:2,自引:0,他引:2  
黄甫  叶盛权 《色谱》2004,22(1):77-80
利用低pH值(pH≤2.0)有效地抑制电渗流,建立起低电渗流毛细管区带电泳(CZE)体系,并分离了7种芳香胺。在此体系中,芳香胺质子化而带正电荷,故采用在毛细管阳极端进样,阴极端检测。实验考察了pH值、电解质浓度对分离的影响,结果发现,当pH<p K a (p K a =14-p K b )时,pH值的微小增大会导致芳香胺的迁移时间迅速延长;芳香胺的出峰次序与其p K b 值及分子中含有的胺基和酸性取代基的数目有关,分子中含胺基愈多,p K b 值愈小,出峰愈早;芳香胺含酸性取代基则使峰序滞后。  相似文献   

15.
傅小芸  吕建德 《分析化学》1992,20(5):524-526
本文采用毛细管电泳法,以50μm内径,45cm长的弹性石英毛细管作为分离管,选用磷酸盐-硼酸盐-十二烷基硫酸钠缓冲溶液体系,在柱254nm紫外检测器,在不同的电泳电压下,对水溶性维生素,磺胺类药物、头孢菌素抗生素,解热镇痛药物有效成份进行了分析,取得较满意的结果。  相似文献   

16.
基于麻黄碱及伪麻黄碱衍生物的光谱及化学性质,设计并构建了毛细管电泳/发光二极管诱导荧光检测系统.对关键光学元件进行组合选择,以蓝光发光二极管为光源,BP 470和BP 530分别为光源滤光片和荧光滤光片,光电倍增管检测信号,并对电泳分离系统的缓冲溶液、分离电压等参数进行优化;以FITC为衍生试剂,10 mmol/L Na2B4O7+ 16 mmol/L SDS为缓冲溶液,12 kV电压下可实现麻黄中麻黄碱和伪麻黄碱的基线分离.在0.25~10 mg/L范围内,麻黄碱和伪麻黄碱标准溶液的质量浓度与荧光响应的峰高之间呈较好的线性关系,相关系数(r)均大于0.99,其检出限分别为0.38 μg/L和0.29 μg/L,峰高的日内重复性(RSD)分别为2.0%和2.2%,日间重复性(RSD)分别为5.4%和5.1%.将该方法用于中药麻黄中麻黄碱和伪麻黄碱的测定,加标回收率分别为94%和107%.  相似文献   

17.
毛细管区带电泳法测定葡萄籽中儿茶素类化合物   总被引:9,自引:0,他引:9  
李奕  高军涛  张志玲  刘虎威  赵保路 《色谱》2000,18(6):491-494
 采用毛细管区带电泳法测定了 10种中国产葡萄籽中的 4个主要儿茶素类化合物 :(+)儿茶素、(- )表儿茶素、(± )表没食子儿茶素、(± )表儿茶素没食子酸酯的含量。在 0 0 2mol/L硼砂和 0 0 0 5mol/L磷酸盐的混合缓冲体系 (pH 10 0 )的背景缓冲液中 ,4个化合物在 10min内取得了令人满意的分离。迁移时间的重现性(RSD)小于 2 % ,峰面积的重现性 (RSD)小于 5 %。在质量浓度为 0 0 0 5g/L~ 0 5 g/L时 ,线性相关系数大于0 995。检测限为 3mg/L~ 10mg/L。该方法简单、快速、准确 ,可作为葡萄籽分析和药用开发过程中分析儿茶素类化合物的有效方法推广使用。  相似文献   

18.
建立了高效毛细管电泳检测盐酸安妥沙星中对映体杂质的方法。以含20 mmol/L HP-β-CD与70mmol/L磷酸二氢钾溶液(用磷酸调节pH3.0)为运行缓冲液,电压30 kV,柱温30℃,检测波长297 nm。可在10 min内将两对映体良好分离,检出限为2 mg.L-1(S/N=3),左旋安妥沙星在4.89~61.2 mg.L-1呈良好的线性关系,日内精密度(RSD)为2.1%(n=5),日间精密度(RSD)为3.2%(n=15)。  相似文献   

19.
毛细管电泳(CE)电导检测(CD)是相对较灵敏和仪器结构简单的一项溶液分析技术,尤其是对于无生色团的无机离子分析更具有突出优势.然而,目前众多商品CE仪器并不配置CD检测器,这极大地限制了该项技术的发展.  相似文献   