首页 | 本学科首页   官方微博 | 高级检索  
     

不同旋转扫描次数对光声层析成像的影响
引用本文:谭毅,李长辉,任亚杰,何军锋. 不同旋转扫描次数对光声层析成像的影响[J]. 应用光学, 2016, 37(5): 700-705. DOI: 10.5768/JAO201637.0502005
作者姓名:谭毅  李长辉  任亚杰  何军锋
作者单位:1.陕西理工大学 物理与电信工程学院,陕西 汉中 723001;
基金项目:国家重大科学仪器专项(2011YQ030114);国家基础研究计划973项目(2011CB707502);国家自然科学基金(11104058)
摘    要:为了实现单元探测器高质量的快速光声图像重建,提出了不同旋转扫描次数对光声层析成像的影响方案。实验采用的光源为YAG激光器,波长为1 064 nm,重复频率为20 Hz,脉宽为7 ns,探测器为针状的PVDF膜水听器,接收直径为1 mm,得到了26个字母、12根头发丝、树叶骨架和模拟血管的光声重建图像。由仿真和实验结果表明,在不牺牲光声重建图像质量的前提下,单元探测器环形扫描一圈,均匀采集100个位置的信号,图像重建时间为5.903 s。该研究结果对于单元探测器的快速旋转扫描成像和环形阵列探测器阵元数的设计具有一定的指导意义。

关 键 词:医用光学与生物技术   光声层析成像   不同旋转扫描次数
收稿时间:2016-03-11

Influence of different rotating scanning times on photoacoustic tomography
Affiliation:1.School of Physics and Telecommunication Engineering,Shaanxi Sci Tech University,Hanzhong 723001,China;2.Department of Biomedical Engineering,College of Engineering,Peking University,Beijing 100871,China
Abstract:In order to perform photoacoustic imaging with high contrast and fast speed, we studied the influence on the reconstructed images of the number of detection positions by a single element transducer circularly scanning targets. A Q switched Nd: YAG laser, operating at 1 064 nm, was used as the illuminating source. The laser has a 7 ns pulse width and a repetition frequency of 20 Hz. A needle polyvinylidene fluoride(PVDF) hydrophone with 1 mm diameter was used to detect photoacoustic signals. We reconstructed the photoacoustic images of 26 letters, 12 hairs, leaf skeleton and simulated blood vessels. Both simulation and experimental results show that an optimized strategy in both imaging quality and speed can be achieved by evenly detecting 100 times around the circle, and the image reconstruction time is 5.903 s. The results of this study would help the performance and design for both scanning single element transducer and ring array systems.
Keywords:
本文献已被 CNKI 等数据库收录!
点击此处可从《应用光学》浏览原始摘要信息
点击此处可从《应用光学》下载免费的PDF全文