The beam-based calibration of an X-ray pinhole camera at SSRF |
| |
Authors: | LENG Yong-Bin, ,HUANG Guo-Qing,ZHANG Man-Zhou,CHEN Zhi-Chu,CHEN Jie,YE Kai-Rong |
| |
Affiliation: | 1,2 1 Shanghai Synchrotron Radiation Facility,Shanghai 201800,China 2 Shanghai Institute of Applied Physics,Shanghai 201800,China |
| |
Abstract: | A pinhole camera for imaging X-ray synchrotron radiation from a dipole magnet is now in operation at the Shanghai Synchrotron Radiation Facility (SSRF) storage ring.The electron beam size is derived by unfolding the radiation image and the point spread function (PSF) with deconvolution techniques.The performance of the pinhole is determined by the accuracy of the PSF measurement.This article will focus on a beam-based calibration scheme to measure the PSF system by varying the beam images with different quadrupole settings and fitting them with the corresponding theoretical beam sizes.Applying this method at SSRF,the PSF value of the pinhole is revised from 37 to 44μm.The deviation in beam size between the theoretical value and the measured value isminimized to 4% after calibration.This optimization allows us to observe the horizontal disturbance due to injection down to as small as 0.5μm. |
| |
Keywords: | beam diagnostics X-ray pinhole camera point spread function beam size measurement |