129I加速器质谱分析研究
引用本文:刘起,周卫健,侯小琳,付云翀,赵稳年,卢雪峰,陈宁,张路远. 129I加速器质谱分析研究[J]. 强激光与粒子束, 2013, 25(8): 2085-2090. DOI: 10.3788/HPLPB20132508.2085
作者姓名:刘起  周卫健  侯小琳  付云翀  赵稳年  卢雪峰  陈宁  张路远
作者单位:1.中国科学院 地球环境研究所, 黄土与第四纪地质国家重点实验室, 西安 71 0075;
摘    要:通过对129I加速器质谱(AMS)分析中影响敏度和准确度各种参数的研究,如靶电极制备、压样、靶样中辅助介质(Matrix)的选择及使用比例等,优化了用于3 MV加速器质谱仪的SO-110型离子源的条件参数,确定129I-AMS测量的最佳靶电极材料为Cu,最佳的辅助介质为Nb粉末,Nb与AgI样品的最佳体积比为3∶1。在此条件下可以获得稳定且持续的I-束流进行测量129I/127I原子比值,实验测得西安加速器质谱仪的129I/127I本底值为1.5210-14。

关 键 词:加速器质谱仪   SO-110离子源   129I   铯溅射   辅助介质
收稿时间:2012-08-02;

Accelerator mass spectrometry analysis of 129I
Liu Qi,Zhou Weijian,Hou Xiaolin,Fu Yunchong,Zhao Wennian,Lu Xuefeng,Chen Ning,Zhang Luyuan. Accelerator mass spectrometry analysis of 129I[J]. High Power Laser and Particle Beams, 2013, 25(8): 2085-2090. DOI: 10.3788/HPLPB20132508.2085
Authors:Liu Qi  Zhou Weijian  Hou Xiaolin  Fu Yunchong  Zhao Wennian  Lu Xuefeng  Chen Ning  Zhang Luyuan
Affiliation:1.State Key Laboratory of Loess and Quaternary Geology,Institute of Earth Environment,Chinese Academy of Sciences,Xi’an 710075; 2.University of Chinese Academy of Sciences,Beijing 100049,China; 3.Shaanxi Key Laboratory of Accelerator Mass Spectrometry Technology and Application,Xi’an AMS Center,Xi’an 710061; 4.Xi’an JiaoTong University,Xi’an 710049,China
Abstract:Ion source is the key part of an Accelerator Mass Spectrometry (AMS) in view of analytical sensitivity, stability and accuracy. SO-110 Cs+ sputtering ion source developed by High Voltage Engineering Europe, is dedicated to AMS for both solid and gas target sources, which has been successfully applied to AMS instruments including one installed in Xi’an AMS center. By investigation of the effect of various parameters, such as the target preparation, holder characteristics, matrix selection and the ratio between AgI and matrix on AMS measurement of 129I using SO-110 ion source, Cu is the suitable holder material, and Nb powder is the best matrix, and the optimal ratio of AgI∶Nb is 1∶3. A machine background of 1.52×10-14 for 129I/127I was obtained using a NaI sample directly mixed with Nb powder and pressed in Cu target holder.
Keywords:SO-110 ion source  129  cesium sputtering  matrix