Concurrent determination of olanzapine,risperidone and 9‐hydroxyrisperidone in human plasma by ultra performance liquid chromatography with diode array detection method: application to pharmacokinetic study |
| |
Authors: | M. Siva Selva Kumar M. Ramanathan |
| |
Affiliation: | 1. Department of Pharmaceutical Analysis, PSG College of Pharmacy, Peelamedu, Coimbatore, Tamilnadu, India;2. Department of Pharmacology, PSG College of Pharmacy, Peelamedu, Coimbatore, Tamilnadu, India |
| |
Abstract: | A simple and sensitive ultra‐performance liquid chromatography (UPLC) method has been developed and validated for simultaneous estimation of olanzapine (OLZ), risperidone (RIS) and 9‐hydroxyrisperidone (9‐OHRIS) in human plasma in vitro. The sample preparation was performed by simple liquid–liquid extraction technique. The analytes were chromatographed on a Waters Acquity H class UPLC system using isocratic mobile phase conditions at a flow rate of 0.3 mL/min and Acquity UPLC BEH shield RP18 column maintained at 40°C. Quantification was performed on a photodiode array detector set at 277 nm and clozapine was used as internal standard (IS). OLZ, RIS, 9‐OHRIS and IS retention times were found to be 0.9, 1.4, .1.8 and 3.1 min, respectively, and the total run time was 4 min. The method was validated for selectivity, specificity, recovery, linearity, accuracy, precision and sample stability. The calibration curve was linear over the concentration range 1–100 ng/mL for OLZ, RIS and 9‐OHRIS. Intra‐ and inter‐day precisions for OLZ, RIS and 9‐OHRIS were found to be good with the coefficient of variation <6.96%, and the accuracy ranging from 97.55 to 105.41%, in human plasma. The validated UPLC method was successfully applied to the pharmacokinetic study of RIS and 9‐OHRIS in human plasma. Copyright © 2015 John Wiley & Sons, Ltd. |
| |
Keywords: | UPLC risperidone 9‐hydroxyrisperidone olanzapine pharmacokinetics |
|
|