首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Ammonium pyrrolidinedithiocarbamate-modified activated carbon micro-column extraction for the determination of As(III) in water by graphite furnace atomic absorption spectrometry
Authors:Yinyan Jiang  Yiwei Wu  Jungfeng Liu  Xinquan Xia  Danlei Wang
Institution:(1) Hubei Key Laboratory of Bioanalytical Technique, Department of Chemistry and Environmental Engineering, Hubei Normal University, Huangshi, P.R. China;(2) Department of Clinical Laboratory, The Second Hospital of Huangshi, Huangshi, P.R. China
Abstract:A simple and selective method using ammonium pyrrolidinedithiocarbamate modified activated carbon (APDC-AC) as solid phase extractant has been developed for speciation of As(III) in water samples. At pH 1.8–3.0, As(III) could be adsorbed quantitatively by APDC-AC, and then eluted completely with 2.0 mL of 0.1 mol L−1 HNO3, while As(V) could almost not be retained at pH 1–7. Effects of acidity, sample flow rate, concentration of elution solution and interfering ions on the recovery of As(III) have been systematically investigated. Under the optimal conditions, the adsorption capacity of APDC-AC for As(III) is 7.3 mg g−1. The detection limit (3σ) of As(III) is 0.05 ng mL−1 for graphite furnace atomic absorption spectrometry (GFAAS) with enrichment factor of 50, and the relative standard deviation (RSD) is 4.1% (n = 9, C = 5 ng mL−1). The method has been applied to the determination of trace As(III) in water, and the recoveries of As(III) are 100 ± 10%. Correspondence: Yiwei Wu, Department of Chemistry and Environmental Engineering, Hubei Normal University, Huangshi 435002, P.R. China
Keywords:: Ammonium pyrrolidinedithio-carbamate modified activated carbon  As(III)  graphite furnace atomic absorption spectrometry            preconcentration and separation
本文献已被 SpringerLink 等数据库收录!